Delta RMC151 User Manual

Page 196

Advertising
background image

RMC70/150 and RMCTools User Manual

b. In one of the task columns, choose the user program you created. Make sure

no other user programs will ever run on that task.

3. Make sure the Task does not stop when an axis halts.

a. In the Project Pane, right-click Programming and choose Properties.
b. On the Halts page, choose Stop the Following Tasks.
c. Uncheck the box for the task that will be running the user program.

Verify the Custom Feedback is Correct
Before verifying that the custom feedback value updates correctly, you will need to make

sure the other feedback values used in the calculation are correct. For example, it your

custom feedback calculates the average of two physical pressure inputs, make sure those

pressure inputs are set up properly and operating correctly before verifying the custom

feedback.
To verify the custom feedback:

a. Make sure the axis is in RUN mode, and the user program you made is running.
b. In the Axis Status Registers, on the All tab, expand the Feedback or

Pressure/Force Feedback section.

c. The Custom Counts register displays the value as written to it by the user program.

Making sure this value updates properly as the axis moves.

Tune the Axis
After completing all the steps above, the axis is ready to be tuned. You will need to

manually tune the axis, because auto-tuning cannot be used in RUN mode.

Error Handling

An axis with custom feedback will receive a No Transducer error if any of the items below

occur. The No Transducer error will then cause the axis to halt according to the No

Transducer Auto Stop setting.

The Custom No Transducer error bit is set.

The Custom Feedback Auto-Fault Mode is violated (see Custom Feedback Auto-

Fault Mode below).

The Custom Counts value is NaN, Inf, or -Inf. These values can result from undefined

operations, such as divide by zero.

176

Delta Computer Systems, Inc.

Advertising
This manual is related to the following products: