Rs-485 communications, Rtu-17 troubleshooting with a digital voltmeter – WattMaster WM-WCC3-TGD-01B User Manual

Page 472

Advertising
background image

10. RS-485 COMMUNICATIONS

WCC III Technical Guide

10-20

RTU-17 Troubleshooting with a Digital Voltmeter

Figure 10-29: Typical RTU-17 RS-485 TUC Communications DC Voltage Measurements with the TUC COMM

Loop Connected – 75176 Driver

Figure 10-30: Typical RTU-17 RS-485 TUC Communications DC Voltage Measurements with the TUC COMM

Loop Connected – LT1785 Driver

The voltage measurements in Figure 10-29 are approximate voltages. Older RTU-17 Controllers used a different RS-485 driver chip.
These voltages are taken when the RTU-17 controller is “ON” and the RS-485 TUC communications loop is connected. These two
voltages will fl uctuate slightly and also will momentarily “fl ip” to the other meter reading’s value. The voltage measurements from “T” to
“SHLD” should be around 2.70 VDC. The voltage measurements from “R” to “SHLD” should be around 3.00 VDC. Typical bad voltages
measurements values would be anything above 3.8 VDC and anything below 1.5 VDC.

The voltage measurements in Figure 10-30 are approximate voltages. Older RTU-17 Controllers used a different RS-485 driver chip.
These voltages are taken when the RTU-17 controller is “ON” and the RS-485 TUC communications loop is connected. These two
voltages will fl uctuate slightly and also will momentarily “fl ip” to the other meter reading’s value. The voltage measurements from “T” to
“SHLD” should be around 2.70 VDC. The voltage measurements from “R” to “SHLD” should be around 3.00 VDC. Typical bad voltages
measurements values would be anything above 3.8 VDC and anything below 1.5 VDC.

Advertising
This manual is related to the following products: