3 meas t/flow - measuring parameters, 4 meas t/flow - evaluations, Measuring parameters – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1023: Evaluation

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1007

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

774 Oven Sample Processor | 874 Oven Sam-
ple Processor

Default value

874 Oven Sample Processor

5.6.4.8.3

MEAS T/Flow - Measuring parameters

Tab: Method

MEAS T/Flow Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measuring time

Maximum period of time for measurement.

Input range

0 - 999999 s

Default value

600 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Input range

1 - 999999.0 s

Default value

10 s

Stop measured value Temp.

Stops when the preset measured value has been reached since the start of
the measurement. off means no stop.

Input range

50.0 - 250.0 °C

Selection

off

Default value

off

5.6.4.8.4

MEAS T/Flow - Evaluations

Tab: Method

MEAS T/Flow Properties... Evaluations

Command name

Name of the command.

Entry

25 characters

The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.