5 det upol - stop conditions, Stop conditions – Metrohm tiamo 2.4 (process analysis) User Manual

Page 569

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

553

Titrando, 855
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware
tab under Sensor to automatic or continuous, then the
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.

Titrando, 855
Input range

-20.0 - 150.0 °C

Default value

25.0 °C

Titrino
Input range

-170.0 - 500.0 °C

Default value

25.0 °C

5.6.3.2.5.5

DET Upol - Stop conditions

Tab: Method

DET Upol Properties... Stop conditions

Command name

Name of the command.

Entry

25 characters

Conditions for stopping the titration. If several stop conditions are set,
then the criterion which is fulfilled first will stop the titration.

Stop volume

Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.

Titrando, 855
Input range

0.00000 - 9999.99 mL

Default value

100,000 mL

Advertising