5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1129

5.6.5.10

CAL LOOP Opt

5.6.5.10.1

CAL LOOP Opt - Overview

Dialog window: Method

CAL LOOP Opt Properties… CAL LOOP Opt - 'Com-

mand name'

Loop command for the calibration of colorimetric sensors, used
together with the CAL MEAS Opt command.

The concentrations which are to be measured with the CAL MEAS Opt
command are defined in this loop command. Each time the loop is exe-
cuted, the validity of the concentration measurement is checked, and at
the end the calibration data for the colorimetric sensor is calculated.

Appearance
The command has the following appearance:

Parameters
The parameters for the CAL LOOP Opt command are set in the following
dialog window:

CAL LOOP Opt - Properties

Command variables
The following command variables are generated by the CAL LOOP pH
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':

Identifica-
tion

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
never been started

.C0

Coefficient c0 calculated from the calibration

.C1

Coefficient c1 calculated from the calibration

.C2

Coefficient c2 calculated from the calibration

.C3

Coefficient c3 calculated from the calibration

.COD

Coefficient of determination R

2

), which is calculated on

the basis of the calibration function and of the confi-
dence interval