5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
862
■■■■■■■■
tiamo 2.4 (for Process Analysis)
■
STAT pH
STAT titration with pH electrodes (measured quantity pH).
■
STAT U
STAT titration with metal electrodes (measured quantity voltage U).
5.6.3.8.2
STAT - Control range
Tab: Method
▶ STAT ▶ Properties... ▶ Control parameters
Set a large control range for steep curves and a small control range for flat
ones. A good approximation for the start of the control range is given by
the point where the tangents intersect.
5.6.3.8.3
STAT - Evaluation window
Dialog window: Method
▶ STAT ▶ Properties... ▶ Evaluations ▶ Rate evalua-
tion
▶ [New]/[Properties] ▶ Evaluation window #
Evaluation windows are defined time segments in which the dosing
rate is determined by linear regression.
Lower limit
Lower limit for the evaluation window.
Input range
0.0 - 99,999.9 s
Default value
0.0 s
Upper limit
Upper limit for the evaluation window.
Input range
0.0 - 999,999.9 s
Default value
999,999.9 s