7 Devices

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1585

Selection

Factory settings | "Method name"

Default value

Factory settings

[Reset calibration]

Resets the device to factory-adjusted calibration. The dark spectrum and
reference spectrum will be deleted.

Determination ID

ID of the determination the wavelength calibration has been carried out
with. With calibration method Factory setting, ---- is displayed.

The determination ID can be copied and pasted into another field.

Calibration coefficients
Coefficients of the regression function with which the assignment of the
CCD pixels to a wavelength is carried out.

c#

Display of the wavelength coefficients c1c4.

7.14.6

Properties - Dark spectrum

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Dark spectrum

The Dark spectrum tab shows information on the dark spectrum stored
in the device.

If no dark spectrum has been recorded yet or the calibration has been
reset to factory settings, -- is displayed in the fields.

Date

Shows the date and time when the dark spectrum was recorded.

User

Shows the user logged in during recording of the dark spectrum.

Method

Shows the method with which the dark spectrum has been recorded.

Determination ID

ID of the determination in which the dark spectrum has been recorded.

The determination ID can be copied and pasted into another field.

Integration time

Shows the integration time during recording of the reference spectrum.