5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

881

Selection

maximum

Default value

maximum

5.6.3.8.6.7

STAT pH - Monitoring

Tab: Method

STAT pH Properties... Monitoring

Command name

Name of the command.

Entry

25 characters

The following monitoring processes can be activated and defined on this
tab:

Monitoring Measured value

Monitoring Measured value

on | off (Default value: off)

If this option is enabled, then the measured value will be monitored and
any violated limits will be entered in the measuring point list.

Lower limit pH

Lower limit of the measured value. If the measured value falls below this
limit, the event Measured value lower limit violated is triggered.

Titrando, 855
Input range

–20.000 - 20.000

Default value

–20.000

Titrino
Input range

–20.00 - 20.00

Default value

–20.00

Lower hysteresis pH

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value again, then the event Measured value
lower limit OK
is triggered.

Titrando, 855
Input range

0.000 - 20.000

Default value

0.020

Upper limit pH

Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit violated is triggered.