5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

936

■■■■■■■■

tiamo 2.4 (for Process Analysis)

STDADD man
Standard addition with manual addition of the standard addition solu-
tion.

STDADD dos
Standard addition with addition of the standard addition solution from
a dosing device.

STDADD auto
Standard addition with automatic addition of the standard addition
solution from a dosing device in such a way that a constant potential
difference results.

5.6.4.2

Standard addition - Overview

Menu item: Method

Insert New command... Measure

Recommendation
In order to guarantee a reliable evaluation of the standard addition, care
must be taken to ensure that the buret volume and the concentration of
the standard solution are adjusted to the respective measuring conditions.
The individual addition volumes must be selected in such a way that the
potential difference is at least 15 mV after each additive step. At least four
volume additions should be carried out.

Table 44

Recommended concentrations of the standard solution for
the standard addition

Buret volume in mL

c

Standard solution

: c

Sample

5

40 : 1

10

20 : 1

20

10 : 1

50

5 : 1

Example of a standard addition

NOTE

This example is valid for standard additions with any type of ion-selec-
tive electrodes.

Sample concentration

5 mg/L

Buret volume

10 mL

Sample size

10 mL

ISA/TISAB

10 mL