2 digital out - general – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1312

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1296

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Once the command has expired, the statuses and the set values of all of
the digital outputs used in the command are saved to command variables.

Parameters
The parameters for the command DIGITAL OUT are configured on the
following 2 tabs:

General

Digital outputs

Command variables
The following command variables are generated in the method run by the
command DIGITAL OUT and can be used in formulas under the designa-
tion 'Command name.Variable designation':

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
never been started

.DEF

Set value which is defined in the configuration for an IO
port; 1 = activated, 0 = deactivated

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command has
never ended

.[VAL]

Value of the command variables (optional, i.e. Command
name.Port name'
= Command name.Port
name.VAL'
). The value returned corresponds to the status
of the IO port after completion of the command; 1 = acti-
vated
, 0 = deactivated

5.6.11.2.2

DIGITAL OUT - General

Tab: Method

DIGITAL OUT General

Command name

Name of the command.

Entry

25 characters

Device

Device name

Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.