6 set ipol - stop conditions, Stop conditions – Metrohm tiamo 2.4 (process analysis) User Manual

Page 770

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

754

■■■■■■■■

tiamo 2.4 (for Process Analysis)

auto
The titration direction is determined automatically from the start mea-
sured value and the set endpoint.

Extraction time

The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). For example, it makes sense to
enter an extraction time for the titration of samples that are not easily
soluble.

Titrando, 719, 720, 736, 751, 758, 785, 794, 798, 799
Input range

0 - 999999 s

Temperature

Manually entered titration temperature. If a temperature sensor is con-
nected and the Temperature measurement is set on the General/
Hardware
tab under Sensor to automatic or continuous, then the
temperature will be measured continuously. This value is used for temper-
ature correction in pH measurements.

Titrando, 855
Input range

-20.0 - 150.0 °C

Default value

25.0 °C

Titrino
Input range

-170.0 - 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando, 855
Input range

0.1 - 999999 s

Default value

2.0 s

Titrino
Input range

0.08 - 16200 s

Default value

2.00 s

5.6.3.4.5.6

SET Ipol - Stop conditions

Tab: Method

SET Ipol Properties... Stop conditions

Command name

Name of the command.

Entry

25 characters