2 cal loop opt - properties – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1146

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1130

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Identifica-
tion

Description

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command
has never ended

.LCO

Loop counter = current number of completed runs, both
of Repeat loops and of While loops

.LST

Start time of the loop command (Date/Time)

.STC{x}

Concentration of the standard x (1 - 50) used for the
command

5.6.5.10.2

CAL LOOP Opt - Properties

Dialog window: Method

CAL LOOP Opt Properties… CAL LOOP Opt - 'Com-

mand name'

The concentrations of the standards, the concentration unit and the type
of the calibration function are defined in the CAL LOOP Opt - 'Com-
mand name'
dialog window.

Command name

Name of the command.

Entry

25 characters

Standards

Unit conc.

Selection of the concentration unit for the standards.

Selection

mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | µg/mL | ppm | % | mEq/L

Default value

ppm

The table must contain at least two but no more than 50 concentrations,
and individual concentration values may appear repeatedly.

Conc. standard

Concentration of the standard.

Input range

0.000 - 999,999.999

Default value

1.000

Moves the concentration upward (modifies sequence).