4 digital out - digital output – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1314

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1298

■■■■■■■■

tiamo 2.4 (for Process Analysis)

If a command block originates from a different device, the Port descrip-
tion
of which does not exist on the currently selected device, then the
affected lines will be presented in red script.

Functions

Moves the selected line up.

Moves the selected line down.

[New]

Opens the dialog window Digital output # in order to edit a new line
(see Chapter 5.6.11.2.4, page 1298).

[Properties]

Opens the dialog window Digital output # in order to edit the selected
line (see Chapter 5.6.11.2.4, page 1298).

[Delete]

Deletes the selected line.

5.6.11.2.4

DIGITAL OUT - Digital output #

Dialog: Method

DIGITAL OUT Digital outputs Properties DIGITAL OUT -

Digital output #

Port description

Selection of the digital output. The digital outputs of the selected device
are available. No port can be selected if not defined is selected in the
Device name field.

Type

Shows the type of the digital output.

Function

Action

Selection

On | Off | Timer | Set default value

Default value

Timer

On
Sets the digital output to on after the start delay has expired.
Off
Sets the digital output to off after the start delay has expired.
Timer
Sets the digital output to on after the Start delay has expired and
then back to off after the Working time has expired.

Advertising