3 meas conc - measuring parameters – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1029

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1013

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Input range

-15 - 15

Default value

8

Switch off automatically

on | off (Default value: off)

If this option is activated, the stirrer will be switched off automatically
when the command has finished.

5.6.4.9.3

MEAS Conc - Measuring parameters

Tab: Method

MEAS Conc Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.