5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1238

■■■■■■■■

tiamo 2.4 (for Process Analysis)

5.6.7.9.2

FLOW - Properties

Dialog window: Method

FLOW Properties... FLOW - 'Command name'

Dialog window: FLOW - 'Command name'

Command name

Name of the command.

Entry

25 characters

Device

Device name

Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Shows the device type. If an instrument is selected in the Device name
field, the Device type field can no longer be edited and the instrument
type belonging to the instrument is displayed.

If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.

Selection

774 Oven Sample Processor | 874 USB Oven
Sample Processor

Default value

874 USB Oven Sample Processor

Gas

Gas type

Gas type of the carrier gas.

Selection

Air | Nitrogen | Other gas

Default value

Air

Gas flow factor

Factor for the correct determination of the gas flow. This parameter is dis-
played only if Gas type = Other gas. For the 774 Oven Sample Pro-
cessor, the gas flow factor values for commonly used carrier gases are lis-