5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

761

[New]

Opens the External measured value # dialog window, in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 934)
.

[Properties]

Opens the External measured value # dialog window, in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 934)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.4.5.10

SET Ipol - Fixed endpoint evaluation

Dialog window: Method

SET Ipol Properties... Additional evalua-

tions

Fixed endpoint evaluation [New/[Properties] Fixed endpoint evalua-

tion #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Input range

-2000.0 - 2000.0 mV

Time
Input range

0.0 - 999999.9 s

Volume
Input range

0.00000 - 9999.99 mL

5.6.3.4.6

SET Upol

5.6.3.4.6.1

SET Upol - Overview

Dialog window: Method

SET Upol Properties... SET Upol - 'Command name'

Command for endpoint titrations with amperometric measurement
(selectable polarization voltage).

Devices
This command can be executed with the following instruments: