2 add - general/hardware – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1176

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1160

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Identifica-
tion

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
never been started

.CONC

Concentration of the solution used for the command

.CYL

Cylinder volume of the exchange or dosing unit used for
the command

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command
has never ended

.TITER

Titer value of the solution used for the command

.VOL

Currently dosed volume

5.6.6.2.2

ADD - General/Hardware

Tab: Method

ADD Properties... General/Hardware

Command name

Name of the command.

Entry

25 characters

The general parameters for the instrument, the dosing device and the stir-
rer are defined on this tab.

Device

Device name

Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead