5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

838

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Delay time

If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time.

Input range

0 - 999 s

Default value

10 s

5.6.3.6.5

KFC - Titration parameters

Tab: Method

KFC Properties... Titration parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the running of the titration are defined on this tab.

Extraction time

The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, e.g. for the titration of sparingly soluble samples.

Input range

0 - 999999 s

Default value

0 s

Temperature

Manually entered titration temperature.

756, 831
Input range

-170.0 - 500.0 °C

Default value

25.0 °C

851, 852
Input range

-20.0 - 150.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

756, 831
Input range

1 - 999999 s

Default value

2 s

851, 852
Input range

0.1 - 999999.0 s

Default value

2.0 s