7 stat u - monitoring, Monitoring – Metrohm tiamo 2.4 (process analysis) User Manual

Page 923

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

907

5.6.3.8.7.7

STAT U - Monitoring

Tab: Method

STAT U Properties... Monitoring

Command name

Name of the command.

Entry

25 characters

The following monitoring processes can be activated and defined on this
tab:

Monitoring Measured value

Monitoring Measured value

on | off (Default value: off)

If this option is enabled, then the measured value will be monitored and
any violated limits will be entered in the measuring point list.

Lower limit

Lower limit of the measured value. If the measured value falls below this
limit, the event Measured value lower limit violated is triggered.

Titrando, 855
Input range

–2,000.0 - 2,000.0 mV

Default value

–2,000.0 mV

Titrino
Input range

–2,000 - 2,000 mV

Default value

–2,000 mV

Lower hysteresis

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value again, then the event Measured value
lower limit OK
is triggered.

Titrando, 855
Input range

0.0 - 2,000.0 mV

Default value

0.2 mV

Upper limit

Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit violated is triggered.

Titrando, 855
Input range

–2,000.0 - 2,000.0 mV

Default value

2,000.0 mV