5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

711

Stop time

Stops when the preset time (including start conditions) has elapsed since
the start of the titration.

Titrando, 855
Input range

0 - 999999 s

Selection

off

Default value

off

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.4.3.7

SET pH - Conditioning

Tab: Method

SET pH Properties... Conditioning

Command name

Name of the command.

Entry

25 characters

Conditioning

on | off (Default value: off)

Enable/disable conditioning. If conditioning is disabled, the following
parameters are not visible.

Start drift

The determination can only be started when the currently measured drift
is less than the entered start drift entered here. If this is the case, then
the status displayed in the Run subwindow in the Workplace program
part changes to COND READY.