2 cal loop ph - properties – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1156

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1140

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Command variables
The following command variables are generated by the CAL LOOP pH
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
never been started

.ENP

Electrode zero point calculated from the calibration

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command has
never ended

.LCO

Loop counter = current number of completed runs, both
of Repeat loops and of While loops

.LST

Start time of the loop command (Date/Time)

.SLO

Electrode slope calculated from the calibration

.VAR

Variance of the calculated result of the calibration in the
selected unit

5.6.5.12.2

CAL LOOP pH - Properties

Dialog window: Method

CAL LOOP pH Properties... CAL LOOP pH - 'Com-

mand name'

The number of buffers and the buffer type are defined in the CAL LOOP
pH - 'Command name'
dialog window.

Command name

Name of the command.

Entry

25 characters

Buffers

Number of buffers

Selection of the number of buffers with which the pH calibration is to be
carried out.

Selection

1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9

Default value

2

Buffer type

Selection of a predefined buffer series for the pH calibration.