8 properties - correction spectrum, Properties - correction spectrum 7, Correction spectrum – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1603

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Devices

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1587

Integration time

Shows the integration time during recording of the reference spectrum.

Averaged spectra

Number of spectra that are recorded and averaged.

Smoothing

Shows the number of neighboring pixels with which the light intensity of
each pixel was determined.

Flash rate

Shows the number of flashes per second of a pulsed xenon lamp as light
source.

Spectrum
Shows the reference spectrum. The graph can be zoomed with the
mouse. You can display the whole spectrum again by double-clicking on
it.

x Axis

Shows the wavelength in nm.

y Axis

Shows the intensity in counts.

7.14.8

Properties - Correction spectrum

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Correction spectrum

As production-related differences with regard to the light sensitivity exist
even among identically built spectrometers, the light intensity of the mea-
suring and reference instrument has to be corrected before a determina-
tion. The resulting correction factors for each wavelength are taken into
account in all subsequent measurements made with the instrument.

The Correction spectrum tab shows information on the correction spec-
trum.

If no correction spectrum has been recorded yet or the calibration has
been reset to factory settings, -- is displayed in the fields.

Date

Shows the date and time when the correction spectrum was recorded.

User

Shows the user logged in during recording of the correction spectrum.