5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

978

■■■■■■■■

tiamo 2.4 (for Process Analysis)

5.6.4.5.6

MEAS Ipol - Fixed endpoint evaluation #

Dialog window: Method

MEAS Ipol Properties... Additional evalua-

tions

Fixed endpoint evaluation [New]/[Property] Fixed endpoint evalua-

tion #

Measured quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value (Titrando, 867, 855)
Input range

-2000.0 - 2000.0 mV

Measured value (Titrino)
Input range

-2000 - 2000 mV

Time
Input range

0.0 - 999999.9 s

5.6.4.6

MEAS Upol

5.6.4.6.1

MEAS Upol - Overview

Dialog window: Method

MEAS Upol Properties... MEAS Upol - 'Command

name'

Command for amperometric measurements with selectable polariza-
tion voltage.

Instruments
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 841, 851, 852, 857, 888, 890, 901, 904,
905, 906, 907

Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798,
799

pH/Ion - Meter: 867

Robotic Titrosampler: 855

Appearance
The command has the following appearance: