7 kft ipol - conditioning, Conditioning – Metrohm tiamo 2.4 (process analysis) User Manual

Page 816

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

800

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Stop time

Stop when the entered time has elapsed after the start of the titration
(including start conditions).

Titrando
Input range

0 - 999,999 s

Selection

off

Default value

off

Filling rate

Rate at which the dosing cylinder is to be filled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing unit used.

Titrando
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.5.2.7

KFT Ipol - Conditioning

Tab: Method

KFT Ipol Properties... Conditioning

Command name

Name of the command.

Entry

25 characters

Conditioning

Conditioning

on | off (Default value: off)

Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.

Start drift

The determination can only be started when the currently measured drift
is lower than the Start drift entered here. In such cases, the Status dis-
played in the Run subwindow on the Single determination or Deter-
mination series
tab will then switch to COND READY.