5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

916

■■■■■■■■

tiamo 2.4 (for Process Analysis)

5.6.3.8.7.10

STAT U - Fixed endpoint evaluation

Dialog window: Method

STAT U Properties... Additional evaluations Fixed

endpoint evaluation

[New/[Properties] Fixed endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Input range

-2000.0 - 2000.0 mV

Time
Input range

0.0 - 999999.9 s

Volume
Input range

0.00000 - 9999.99 mL

5.6.3.9

TET

5.6.3.9.1

TET - Overview

Dialog window: Method

TET Properties... TET - 'Command name'

Command for thermometric titration with the 859 Titrotherm.

Instruments
This command can be executed with the following device:

Titrando: 859

Appearance
The command has the following appearance:

Parameters
The parameters for the command TET are configured on the following
tabs: