8 kft ipol - additional evaluations, Additional evaluations – Metrohm tiamo 2.4 (process analysis) User Manual

Page 819

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

803

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the
subwindow Live display of the program part Workplace, in addition to
the drift and the time.

5.6.3.5.2.8

KFT Ipol - Additional evaluations

Tab: Method

KFT Ipol Properties... Additional evaluations

Command name

Name of the command.

Entry

25 characters

The following additional methods for evaluation of titration curves can be
activated and defined on this tab:

Fixed endpoint evaluation

Fixed endpoint evaluation

on | off (Default value: off)

If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Volume or Time) for the fixed endpoint from the
measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the Fixed endpoint evaluation # dialog window for entering a
new fixed endpoint (see Chapter 5.6.3.5.2.10, page 806).

[Properties]

Opens the Fixed endpoint evaluation # dialog window to edit the
fixed endpoint selected in the table (see Chapter 5.6.3.5.2.10, page
806)
.

[Delete]

Deletes the selected line.