5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

854

■■■■■■■■

tiamo 2.4 (for Process Analysis)

5.6.3.7.5

BRC - Titration parameters

Tab: Method

BRC Properties... Titration parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the running of the titration are defined on this tab.

Extraction time

The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, e.g. for the titration of sparingly soluble samples.

Input range

0 - 999999 s

Default value

0 s

Temperature

Manually entered titration temperature.

Input range

-20.0 - 150.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Input range

0.1 - 999999.0 s

Default value

2.0 s

5.6.3.7.6

BRC - Stop conditions

Tab: Method

BRC Properties... Stop conditions

Command name

Name of the command.

Entry

25 characters

The conditions for stopping the titration are defined on this tab.

Stop time

Stop when the entered time has elapsed since the start of the titration.
off means no stop.

Input range

1 - 999999 s

Selection

off

Default value

off