5 calibration commands, 1 calibration commands - overview, Calibration commands 5 – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1111

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1095

5.6.5

Calibration commands

5.6.5.1

Calibration commands - Overview

Menu item: Method

Insert New command... Calibration

Commands for calibration of pH electrodes and ion-selective electrodes.

General
The commands CAL LOOP and CAL MEAS are available for calibrating
electrodes. The number and type of buffers or standards, respectively, that
can be measured with the CAL MEAS command is defined in the CAL
LOOP
loop command. Each time the loop is executed, the validity of the
measurement is checked, and at the end the calibration data for the elec-
trode is calculated.

The basic structure of a calibration method looks as follows: