7 Devices

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1583

Entry

1000 characters

Default value

empty

7.14.3

Properties - Settings

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Settings

Device assignment

Instrument for reference channel

Name of the reference instrument assigned to the instrument for the mea-
suring channel in a two-channel instrument setup. It measures the refer-
ence, while the instrument for the measuring channel measures the sam-
ple solution at the same time.

Selection

'Instrument name' | not defined

'Instrument name'
All spectrometers for which Instrument for reference channel =
not defined is selected.
not defined
In a one-channel instrument setup and for spectrometers that have
already been assigned in a two-channel instrument setup, not
defined
has to be selected.

Instrument for measuring channel

Only visible for two-channel instrument setups. Name of the instrument
for sample solution measurement.

Light source

Type

Type of light source used.

Selection

continuous | pulsed

Default value

continuous

Analog output

on | off (Default value: on)

If this option is enabled, then the measuring signal is output at the analog
output of the spectrometer. If the connected spectrometer is not
equipped with an analog output, then this option will be ignored.

Sensitivity

Setting of the sensitivity of the analog output. The number of mAU
(absorption units) represented by 1 mV is indicated.