5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

868

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Identification

Description

.RE{x}.RWL

Lower limit of the evaluation window × (1 - 9) in s

.RE{x}.RWH

Upper limit of the evaluation window × (1 - 9) in s

.SLO

Electrode slope of the sensor used for the command
(in %)

.SME

Start measured value (measured value after process-
ing the start conditions) in the unit of the measured
value

.STE

Start temperature (temperature after processing the
start conditions) in °C

.STY

Type of stop with which the command was stopped:
1 = normal; 0 = manual or after error

.SVA

Start volume absolute (volume that was added
according to the start condition "start volume") in mL

.SVT

Total start volume (volume that was added according
to all three start conditions) in mL

.TITER

Titer value of the solution used for the command

5.6.3.8.6.2

STAT pH - General/Hardware

Tab: Method

STAT pH Properties... General/Hardware

Command name

Name of the command.

Entry

25 characters

The general parameters for the instrument, the dosing device, the sensor
and the stirrer are defined on this tab.

Device

Device name

Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.