3 properties - rs-232, Properties - rs-232 7 – Metrohm tiamo 2.4 (process analysis) User Manual
Page 1613
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
tiamo 2.4 (for Process Analysis)
■■■■■■■■
1597
Entry
1000 characters
Default value
empty
7.16.3
Properties - RS-232
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ RS-232
The properties of the RS-232 interface are defined on this tab.
COM Port
Selection of the serial interface on the PC to which the device is con-
nected.
Selection
COM1 | COM2 | … | COMn
Default value
COMn
COMn
First free COM port.
Baud rate
Transmission rate. The baud rate selected here must also be set on the
device itself.
Selection
300 | 600 | 1200 | 2400 | 4800 | 9600 | 19200 |
38400 | 57600 | 115200
Default value
9600
Data bit
Number of data bits.
Selection
7 | 8
Default value
8
Parity
Type of parity testing.
Selection
None | Odd | Even
Default value
None
Stop bit
Number of stop bits.
Selection
1 | 2
Default value
1
Handshake
Type of data transmission protocol.
Selection
None | HW | SW
Default value
None