2 loop - properties – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1305

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1289

(case 1), the loop is run through. It is possible with certain stop conditions
that the loop is never run through.

NOTE

Nested and overlapping loops are not allowed.

Parameters
The parameters for the command LOOP are configured in the following
dialog window:

LOOP

Command variables
The following command variables are generated in the method run by the
command LOOP and can be used in formulas under the designation
'Command name.Variable designation':

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

.LCO

Loop counter = current quantity of completed runs, both
of Repeat loops and of While loops

.LST

Start time of the loop command (Date/Time)

5.6.10.4.2

LOOP - Properties

Dialog window: Method

LOOP Properties... LOOP - Loop #

Command name

Name of the command.

Entry

25 characters

Stop conditions
Definition of three possible stop conditions. The first stop condition to be
fulfilled will stop the loop.

Advertising