5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1113

The reference wavelengths with which the spectrometer is calibrated can
be entered on this tab.

Tolerance

Tolerance for the preset wavelength when assigning the highest peak
maximum found during the calibration measurement to the reference
wavelength. If a peak is outside the range Wavelength +/- Tolerance,
the calibration will be canceled. The existing calibration in the configura-
tion remains in effect.

Input range

0.1 - 1,100.0 nm

Default value

0.5 nm

[New]

Opens the Calibration wavelength # dialog window for entering wave-
lengths and tolerances (see Chapter 5.6.5.5.5, page 1113).

[Properties]

Opens the Calibration wavelength # dialog window to edit the values
selected in the table (see Chapter 5.6.5.5.5, page 1113).

[Delete]

Deletes the selected line.

5.6.5.5.5

CAL Spec - Calibration wavelength #

Tab: Method

CAL Spec Properties... Calibration wavelengths [New/[Prop-

erties]

Calibration wavelength #

Calibration wavelengths

Input range

200.0 - 1,100.0 nm

Default value

'empty' nm

5.6.5.6

CAL LOOP Conc

5.6.5.6.1

CAL LOOP Conc - Overview

Dialog window: Method

CAL LOOP Conc Properties... CAL LOOP Conc -

'Command name'

Loop command for the calibration of ion-selective electrodes, used
together with the CAL MEAS Conc command.