8 stat u - evaluations, Evaluations – Metrohm tiamo 2.4 (process analysis) User Manual

Page 928

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

912

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Selection

Cancel determination | Cancel command
(Titrando, 855)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT U command will be canceled, then the exit track (if
present) will be started and the determination will be finished.
Cancel command (Titrando, 855)
The running STAT U command will be canceled, then the next com-
mand will be carried out.
Wait for [Continue]
Reagent dosing in the running STAT U command will be interrupted
and a message will be displayed. As soon as the monitored measured
value is again within the limits (including hysteresis), reagent dosing
can be resumed by pressing [Continue] in this message box.
Wait for limit ok
Reagent dosing in the running STAT U command will be interrupted.
As soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are violated.

Track call on limit exceeding
In this table, which cannot be edited directly, a maximum of 20 entries
can be defined as to which track is to be started automatically when a
particular limit value is violated.

[New]

Opens the Track call # dialog window, in which the parameters for the
track call can be entered (see Chapter 5.6.3.8.4, page 863).

[Properties]

Opens the Track call # dialog window, in which the parameters for the
track call can be edited (see Chapter 5.6.3.8.4, page 863).

[Delete]

Deletes the track call selected in the table.

5.6.3.8.7.8

STAT U - Evaluations

Tab: Method

STAT U Properties... Evaluations

Command name

Name of the command.

Entry

25 characters

Advertising