5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

565

greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.

5.6.3.2.5.11

DET Upol - Fixed endpoint evaluation

Dialog window: Method

DET Upol Properties... Additional evalua-

tions

Fixed endpoint evaluation [New/[Properties] Fixed endpoint evalua-

tion #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Input range

–200.0 - 200.0 µA

Time
Input range

0.0 - 999,999.9 s

Volume
Input range

0.00000 - 9,999.99 mL

5.6.3.3

MET

5.6.3.3.1

MET - Overview

Menu item: Method

Insert New command...

Command for Monotonic equivalence point titrations.