8.4 Manual control - Functions

■■■■■■■■■■■■■■■■■■■■■■

1686

■■■■■■■■

tiamo 2.4 (for Process Analysis)

External position

Selection of the external position to which the current robotic arm posi-
tion is to be assigned.

Input range

1 - 4

Default value

1

[Assign]

Triggers assignment. During the assignment the cursor is displayed as
hourglass.

8.4.4.5

Sample changer - Pumping

Program part: Manual

The pumps connected to the tower can be switched on and off here.

Pump 1
Manually controlling pump 1 or valve 1 on the selected tower.

Operation

Selection whether the pump or valve is to be switched on or off manually
or whether it is to be switched on for a specified duration and then
switched off again automatically.

Selection

On/Off | Duration

Default value

On/Off

Duration

Entry of the duration during which the pump/valve should remain
switched on. This field is only displayed if Operation = Duration.

Input range

1 - 999,999 s

Default value

60 s

[Start]

Starts pump 1. The time elapsed since the start is shown in the status dis-
play.

[Stop]

Stops pump 1.

Pump 2
Manually controlling pump 2 or valve 2 on the selected tower.

Operation

Selection whether the pump or valve is to be switched on or off manually
or whether it is to be switched on for a specified duration and then
switched off again automatically.