5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1204

■■■■■■■■

tiamo 2.4 (for Process Analysis)

the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Entry

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

5.6.6.4.3

LQH - Parameters

Tab: Method

LQH Properties... Parameters

Command name

Name of the command.

Entry

25 characters

The liquid handling function is defined on this tab.

Function

Selection of the liquid handling function that is to be carried out.

Selection

Dosing | Fill | Aspirate | Eject to stop | Eject to
end volume
| Exchange position | Change
port
| Compensate

Default value

Dosing

Dosing
The given volume is added. No automatic filling takes place either
before or after the command.
Fill
Fill
the cylinder can take place from a freely selectable port. The valve
disc then remains at the selected port.
Aspirate
With this function, liquid is aspirated from the given port. As with Dos-
ing, the cylinder is not filled automatically beforehand or afterwards. It
should be possible to achieve the volume to be aspirated with a single
piston stroke.
Eject to stop
With this function, the entire cylinder content is ejected via the speci-
fied port. The piston is lowered to the stop point, i.e. past the maxi-
mum volume mark. This function should only be carried out to eject
any air bubbles.