5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1042

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Identification

Description

.STY

Type of stop with which the command was stopped:
1 = normal; 0 = manual or after error

.PK{#}.ABS

Absorbance of the peak with the greatest absorbance
(maximum absorbance) for evaluation # (1 - 9)

.PK{#}.TRN

Transmission of the peak with the greatest absorb-
ance (minimum transmission) for evaluation # (1- 9) in
%

.PK{#}.CNT

Intensity of the peak with the greatest absorbance
(maximum intensity) for evaluation # (1 - 9)

.PK{#}.CNR

Intensity in reference spectrum 'Command
name'.PK{#}.WVL for evaluation # (1 - 9)

.PK{#}.CND

Intensity in dark spectrum 'Command
name'.PK{#}.WVL for evaluation # (1 - 9)

.PK{#}.WVL

Wavelength of the peak with the greatest absorbance
for evaluation # (1 - 9) of the discovered maximum

.PK{#}.SAT

Specification as to whether the detector was satura-
ted with 'Command name'.PK{#}.WVL for evaluation
# (1 - 9)

5.6.4.13.2

MEAS Spec - General/Hardware

Tab: Method

MEAS Spec Properties... General/Hardware

Command name

Name of the command.

Entry

25 characters

The general parameters for the instrument are defined on this tab.

Device

Device name

Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.