5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

819

Titrando
Input range

1 - 999 µL/min

Default value

10 µL/min

Delay time

If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time. The selection inf. stands for infin-
ite.

Titrando, 855
Input range

0 - 999 s

Default value

10 s

Titrino
Input range

0 - 999 s

Default value

10 s

Selection

inf.

Stop time

The titration is canceled when the absolute stop time defined here has
elapsed since the start of the titration. This parameter is displayed only if
Stop criterion = Time and if Delay time = inf..

Titrino
Input range

1 - 999,999 s

5.6.3.5.3.5

KFT Upol - Titration parameters

Tab: Method

KFT Upol Properties... Titration parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the running of the titration are defined on this tab.

Titration direction

Selection of the titration direction.

Selection

+ | - | auto

Default value

auto

+
positive measured value change
-
negative measured value change