11 met ipol - fixed endpoint evaluation – Metrohm tiamo 2.4 (process analysis) User Manual

Page 662

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

646

■■■■■■■■

tiamo 2.4 (for Process Analysis)

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.

5.6.3.3.4.11

MET Ipol - Fixed endpoint evaluation

Dialog window: Method

MET Ipol Properties... Additional evalua-

tions

Fixed endpoint evaluation [New]/[Properties] Fixed endpoint evalua-

tion #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value
Input range

-2000.0 - 2000.0 mV

Time
Input range

0.0 - 999999.9 s

Volume
Input range

0.00000 - 9999.99 mL

Advertising