5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1225

NOTE

Please make sure that none of the lift positions exceed the maximum
stroke path given in the tower properties of the device.

Parameters

Lift rate

Selection of the speed for moving the lift.

778, 789, 814, 815, 855, 864, 874
Input range

5 - 25 mm/s

Default value

25 mm/s

730
Input range

3 - 25 mm/s

Default value

25 mm/s

774
Input range

3 - 12 mm/s

Default value

12 mm/s

5.6.7.5

PUMP

5.6.7.5.1

PUMP - Overview

Dialog window: Method

PUMP Properties... PUMP - 'Command name'

Command for controlling pumps connected to or built into the Sample
Processor.

Devices
This command can be executed with the following devices:

Sample Processor: 730, 778, 789, 814, 815, 855, 864

Appearance
The command has the following appearance:

Parameters
The parameters for the command PUMP are configured in the following
dialog window:

PUMP - Properties