3 excellent electrode – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1385

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1369

6.3.3.6.3

Excellent electrode

Tab: Configuration

Tools Templates Electrode types... [Proper-

ties]

Excellent electrode

Max. streaming potential

Maximum permitted flow potential (difference of the measured voltages
in stirred and non-stirred solution).

Input range

0.0 - 999.9 mV

Default value

2.5 mV

Max. drift

Maximum permitted drift in stirred solution.

Input range

0.1 - 9.9 mV/min

Default value

2.0 mV/min

Min. slope

Minimum permitted slope of the pH electrode.

Input range

-999.9 - 999.9 %

Default value

96.5 %

Max. slope

Maximum permitted slope of the pH electrode.

Input range

-999.9 - 999.9 %

Default value

101.0 %

Max. response time

Maximum permitted response time of the pH electrode. The voltage mea-
sured after three minutes in stirred solution is used as a comparison value
for determining the response time. The response time is the time after
which the measured voltage has reached this comparison value except for
±1 mV.

Input range

0 - 999 s

Default value

45 s

Min. offset potential

Minimum permitted offset voltage (voltage at pH = 7.0). This value applies
for all evaluations.

Input range

-999 - 999 mV

Default value

-20 (Gel), -50 (Non-aqueous), -15 (Stand-
ard) mV