5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

847

Designation

Description

.LP.IGE

Current pulsed current for the last measuring point in
the measuring point list in mA

.LP.MEA

Measured value for the last measuring point in the
measuring point list in mV

.LP.QTY

Measured value (water) for the last measuring point in
the measuring point list in µg

.LP.TEM

Temperature for the last measuring point of the mea-
suring point list in °C

.LP.TIM

Time in s until the last measuring point in the measur-
ing point list is reached

.LP.UGE

Voltage at the generator electrode for last measuring
point of the measuring point list in V

.MTE

Temperature measurement with sensor; 1 = on, 0 =
off

.NMP

Number of measuring points in measuring point list

.SME

Start measured value (measured value after processing
the start conditions) in mV

.STE

Start temperature (temperature after processing of the
start conditions) in °C

5.6.3.7.2

BRC - General/Hardware

Tab: Method

BRC Properties... General/Hardware

Command name

Name of the command.

Entry

25 characters

The general parameters for the device, the electrodes and the stirrer are
defined on this tab.

Device

Device name

Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.

Selection

'Device name' | not defined

Default value

not defined