tiamo 2.4 (for Process Analysis)

Only editable if the analog output is switched on.
Input range

1 - 214,783,647 mAU/mV

Default value

100 mAU/mV

7.14.4

Properties - Detector

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Detector

The Detector tab shows information on the detector that is built into the
spectrometer.

Detector type

Shows the detector type.

Minimum wavelength

Shows the minimum wavelength of the measurable range in nm (instru-
ment-specific parameter).

Maximum wavelength

Shows the maximum wavelength of the measurable range in nm (instru-
ment-specific parameter).

Number of pixels

Shows the resolution with which the spectrum is recorded (instrument-
specific parameter).

7.14.5

Properties - Calibration

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Calibration

On the tab Calibration the data of the wavelength calibration is dis-
played. A specific wavelength, defined by four wavelength coefficients, is
assigned to each pixel on the detector. The wavelength calibration is car-
ried out with the command CAL Spec, the calibration data is calculated
automatically and displayed on this tab.

Date

Display of the time and date of the wavelength calibration.

User

Display of the user logged in at the time of the wavelength calibration.

Method

Display of the method the wavelength calibration has been carried out
with.