2 meas conc - general/hardware – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1027

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1011

Designa-
tion

Description

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command
has never ended

.IME

Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value

.ITE

Initial temperature (temperature before start conditions
are processed) in °C

.LP.CAx

Calculated value × (1...3) for the last measuring point on
the measuring point list

.LP.EXx

External value × (1...3) for the last measuring point of
the measuring point list

.LP.MEA

Measured value for the last measuring point of the mea-
suring point list in the unit of the measured value

.LP.TEM

Temperature for the last measuring point of the measur-
ing point list in °C

.LP.TIM

Time in s until the last measuring point in the measuring
point list is reached

.MTE

Temperature measurement with sensor; 1 = on, 0 =
off
)

.NMP

Number of measuring points in measuring point list

.SLO

Electrode slope of the sensor used for the command (in
mV for ISE sensors)

.STY

Type of stop with which the command was stopped: 1
= normal
; 0 = manual or after error

5.6.4.9.2

MEAS Conc - General/Hardware

Tab: Method

MEAS Conc Properties... General/Hardware

Command name

Name of the command.

Entry

25 characters

The general parameters for the device, the sensor and the stirrer are
defined on this tab.