5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1266

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Command variables
The following command variables are generated in the method run by the
command SEND and can be used in formulas under the designation
'Command name.Variable designation':

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

5.6.9.4.2

SEND - Properties

Dialog window: Method

SEND Properties... SEND - 'Command name'

Command name

Name of the command.

Entry

25 characters

All the Event messages listed in the table are sent simultaneously. The
table cannot be edited directly.

NOTE

A maximum of 10 event messages can be entered.

[New]

Opens the dialog window Send event message for entering a new
event message (see Chapter 5.6.9.4.3, page 1267).

[Properties]

Opens the dialog window Send event message to edit the event mes-
sage selected in the table (see Chapter 5.6.9.4.3, page 1267).

[Delete]

Deletes the selected event message.