4 options, 1 options - overview, Options 3 – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1389: General program properties

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1373

Electrode type "Gel"

Parameters

Excellent

electrode

Good

electrode

Usable

electrode

Max. streaming potential
(mV)

3.0

3.5

4.5

Max. drift (mV/min)

2.5

3.0

4.0

Min. slope (%)

96.5

96.0

95.0

Max. slope (%)

101.0

102.0

103.0

Max. response time (s)

60

75

90

Min. offset potential (mV)

-20

-20

-20

Max. offset potential (mV)

20

20

20

Electrode type "Non-aqueous"

Parameters

Excellent

electrode

Good

electrode

Usable

electrode

Max. streaming potential
(mV)

3.0

4.5

6.0

Max. drift (mV/min)

5.0

7.0

9.0

Min. slope (%)

88.0

80.0

70.0

Max. slope (%)

120.0

130.0

140.0

Max. response time (s)

60

75

90

Min. offset potential (mV)

-50

-50

-50

Max. offset potential (mV)

100

100

100