5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1202

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Appearance
The command has the following appearance:

Parameters
The parameters for the command LQH are configured on the following 2
tabs:

General/Hardware
Parameters for device and dosing device.

Parameters
Parameters for adjusting the Liquid Handling function.

Command variables
The following command variables are generated in the method run by the
command LQH and can be used in formulas under the designation 'Com-
mand name.Variable designation'
:

Identifica-
tion

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.CONC

Concentration of the solution used for the command

.CYL

Cylinder volume of the exchange or dosing unit used for
the command

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

.TITER

Titer value of the solution used for the command

.VOL

Currently dosed volume

5.6.6.4.2

LQH - General/Hardware

Tab: Method

LQH Properties... General/Hardware

Command name

Name of the command.

Entry

25 characters