6 det ph - potentiometric evaluation, Potentiometric evaluation – Metrohm tiamo 2.4 (process analysis) User Manual

Page 490

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

474

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Volume after EP

When the number of equivalence points defined under Stop EP has been
found, this volume will be added. In this way you can see the curve shape
after the equivalence point is reached.

Titrando, 855
Input range

0.01000 - 9999.99 mL

Selection

off

Default value

off

Stop time

Stops when the preset time (including start conditions) has elapsed since
the start of the titration.

Titrando, 855
Input range

0 - 999999 s

Selection

off

Default value

off

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.

Titrando, 855
Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Input range

0.01 - 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.2.2.6

DET pH - Potentiometric evaluation

Tab: Method

DET pH Properties... Potentiometric evaluation

Command name

Name of the command.

Entry

25 characters

Parameters for the potentiometric evaluation of the titration curve with
automatic equivalence point recognition. One of the following evaluation
methods can be chosen: