4 stat ph - control parameters, Control parameters – Metrohm tiamo 2.4 (process analysis) User Manual
Page 891
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
875
5.6.3.8.6.4
STAT pH - Control parameters
Tab: Method
▶ STAT pH ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Entry
25 characters
Control point
Control point at pH
Definition of the pH value to be controlled.
Titrando, 855
Input range
–20.000 - 20.000
Selection
off
Default value
off
Titrino
Input range
–20.00 - 20.00
Selection
off
Default value
off
Titration rate
Titration rate
Three predefined rates, for which the parameters for Control are not dis-
played, can be selected for the titration rate. The Users setting must be
selected in order to be able to edit these parameters.
Selection
50 µL/min | 100 µL/min | 500 µL/min | Users
Default value
100 µL/min
Control
NOTE
Will be displayed only if the Users option is selected in the selection list
Titration rate.
Dynamics pH
The control range defines the measured value range before the specified
control point. It has a decisive influence on the titration rate and thus also
on the accuracy (see Chapter 5.6.3.8.1, page 860).